TopPette Single Channel Manual Pipette

TopPette
Pipettes are suitable for precise sampling and transfer of liquids. Using the principle of air displacement, the liquid is transferred using a removable disposable nozzle. The pipette is composed of a positioning part, a volume adjustment indicating part, an ejector device, a piston sleeve and a suction nozzle.

FEATURES

•   Lightweight, ergonomic, low force design

•   Digital display clearly reads volume setting

•   The pipettes cover volume range of 0.1μl to 10ml

•   Easy to calibrate and maintain with tool supplied

•   Design helps avoid repetitive strain injuries

•   Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate

•   The low part is available for autoclaving

SPECIFICATIONS


Single-channel Adjustable Volume   Pipettes

Cat. No.

Volume Range

Increment

Test Volume

Accuracy error

Precision error





%

μl

%

μl

7010101001

0.1-2.5μl

0.05μl

2.5

2.50%

0.0625

2.00%

0.05

1.25

3.00%

0.0375

3.00%

0.0375

0.25

12.00%

0.03

6.00%

0.015

7010101004

0.5-10μl

0.1μl

10

1.00%

0.1

0.80%

0.08

5

1.50%

0.075

1.50%

0.075

1

2.50%

0.025

1.50%

0.015

7010101005

2-20μl

0.5μl

20

0.90%

0.18

0.40%

0.08

10

1.20%

0.12

1.00%

0.1

2

3.00%

0.06

2.00%

0.04

7010101006

5-50μl

0.5μl

50

0.60%

0.3

0.30%

0.15

25

0.90%

0.225

0.60%

0.15

5

2.00%

0.1

2.00%

0.1

7010101008

10-100μl

1μl

100

0.80%

0.8

0.15%

0.15

50

1.00%

0.5

0.40%

0.2

10

3.00%

0.3

1.50%

0.15

7010101009

20-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

20

3.00%

0.6

1.00%

0.2

7010101011

50-200μl

1μl

200

0.60%

1.2

0.15%

0.3

100

0.80%

0.8

0.30%

0.3

50

1.00%

0.5

0.40%

0.2

7010101014

100-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

100

2.00%

2

0.70%

0.7

7010101016

200-1000μl

5μl

1000

0.60%

6

0.20%

2

500

0.70%

3.5

0.25%

1.25

200

0.90%

1.8

0.30%

0.6

7010101017

1000-5000μl

50μl

5000

0.50%

25

0.15%

7.5

2500

0.60%

15

0.30%

7.5

1000

0.70%

7

0.30%

3

7010101033

2-10ml

0.1ml

10ml

0.60%

60

0.20%

20

5ml

1.20%

60

0.30%

15

2ml

3.00%

60

0.60%

12



Fixed Volume Pipettes

Cat. No.

Volume Range

Increment

Test Volume

Accuracy error

Precision error





%

μl

%

μl

7010102018

5μl

-

5μl

1.30%

0.065

1.20%

0.06

7010102019

10μl

-

10μl

0.80%

0.08

0.80%

0.08

7010102020

20μl

-

20μl

0.60%

0.12

0.50%

0.1

7010102021

25μl

-

25μl

0.50%

0.125

0.30%

0.075

7010102022

50μl

-

50μl

0.50%

0.25

0.30%

0.15

7010102023

100μl

-

100μl

0.50%

0.5

0.30%

0.3

7010102024

200μl

-

200μl

0.40%

0.8

0.20%

0.4

7010102025

250μl

--

250μl

0.40%

1

0.20%

0.5

7010102026

500μl

-

500μl

0.30%

1.5

0.20%

1

7010102027

1000μl

-

1000μl

0.30%

3

0.20%

2

7010102028

2000μl

-

2000μl

0.30%

6

0.15%

3

7010102029

5000μl

-

5000μl

0.30%

15

0.15%

7.5

REQUEST A QUOTE